My ClickBD
DY21 single channel syringe pump features: 1) 3.5 inches LCD screen for clear vision within 5 meters. 2) Special designed extension line grip. 3) One-button night mode design for patient's comfortable at night. 4) Compatible with different bands of 5, 10, 20/30, 50/60 ml syringes. 5) Free stacking of single syringe pumps for easy space station in ICU, NICU, OT etc. Detailed description: Channel: Single. Infusion Modes: Rate Mode, Rate-Time, Rate-VTBI, Time-VTBI, Body weight. Safety: Dual CPU to make sure higher standard infusion. Drug Library: 20 drug list with drug code display. Accuracy: More or less 2%. Time Preset: 00:01~99:59(Hour:Min). Volume Range: 0~9999.9 ml. Flow Rate Step by: 0.1 ml/h when Rate=100 ml/h. Flow Rate Range: For 5ml syringe, 0.1ml/h~150ml/h; For 10ml syringe, 0.1ml/h~300ml/h; For 20ml syringe, 0.1ml/h~600ml/h; For 30ml syringe, 0.1ml/h~900ml/h; For 50-60ml syringe, 0.1ml/h~1200ml/h. Purge/Bolus Rate/Top Flow Rate: 5ml syringe-150ml/h; 10ml syringe-300ml/h; 20ml syringe-600ml/h; 30ml syringe-900ml/h; 50/60ml syringe-1200ml/h. Audible & Visible Alarms: Auto Self-Test, Syringe Dislocation, Occlusion, Near End, Syringe Empty, VTBI Completion, Low Battery, Battery Exhausted, Motor Malfunction, Incorrect Syringe Specification, Circuit Malfunction, Master CPU Malfunction, Monitor CPU Malfunction, Parameters Over Limitation, A.C. Connection/Disconnection. KVO: 0.1~5.0ml/h Adjustable. Occlusion Pressure: High: 40KPa +- 20KPa; Medium: 60KPa +- 20KPa; Low: 100KPa +-20KPa. Battery: >=10 hours. Power Consumption: 30VA. Power Supply: AC: 100-240V;50Hz/60Hz. Classification: Class I; Type CF; IPX4. Weight: 2 kg. Optional Function: Ambulance DC 12V Power. |